Emergency numbers differ by country. This page uses 911, the number for the United States and Canada. Many other countries use 112, 999, or another number. Find the number for your country.
Natural disasters, severe winter storms, wildfires, and extended electrical grid collapses disproportionately endanger individuals with mobility disabilities, sensory impairments, neurodivergent conditions, and chronic medical dependencies. When an emergency strikes, standard public infrastructure frequently fails: building elevators ground and lock out, cellular networks saturate, accessible paratransit halts, and retail pharmacies close.
True disaster resilience for medical and functional needs requires engineering independent redundancy before a crisis begins. This guide provides watt-hour battery calculations for life-sustaining Durable Medical Equipment (DME), compressed oxygen tank duration tables, cold-chain insulin preservation rules, multi-story stair evacuation mechanics, and legal shelter rights aligned with the US Federal Emergency Management Agency (FEMA), the US Department of Health and Human Services (HHS emPOWER), the Americans with Disabilities Act (ADA), the World Health Organization (WHO), and the United Nations Office for Disaster Risk Reduction (UNDRR).
ھاڻي ھي ڪر¶
- If a life-sustaining medical device, such as a mechanical ventilator, left ventricular assist device (LVAD), or high-flow oxygen system, has less than 2 to 4 hours of battery or tank reserve left during a blackout, call your local ايمرجنسي نمبر فوري طور تي (آمريڪا ۽ ڪئناڊا ۾ 911، يورپي يونين ۾ 112، برطانيه ۾ 999) هڪ متحرڪ سهولت لاء طبي ٽرانسپورٽ لاء.
- Switch all positive airway pressure (CPAP / BiPAP) and ventilator units to direct current (12V / 24V DC) battery cables and turn off heated humidifiers and heated tubing, which immediately reduces battery drain by 65 to 75 percent.
- جيڪڏهن توهان هڪ اسٽيشنري اليڪٽرڪ آڪسيجن ڪنسنٽريٽر استعمال ڪريو ٿا، ته فوري طور تي پنهنجي بيڪ اپ ڪمپريسڊ گيس آڪسيجن سلنرز (اي-ٽينڪ يا M60 ٽينڪ) يا بيٽري سان هلندڙ پورٽيبل آڪسيجن ڪنسنٽريٽر کي پلس-ڊوز موڊ تي سيٽ ڪيو، ۽ تصديق ڪريو ته پيتل جي ٽانڪي رينچ کي جسماني طور تي ٻيهر گل ڪرڻ لاءِ.
- Protect refrigerated medications, including insulin, biologics, and liquid antibiotics, inside an insulated cooler using chilled gel packs separated from the medicine vials by a folded towel so the medication never freezes solid.
- توهان جي ٽن ماڻهن جي پرسنل سپورٽ نيٽ ورڪ کي خبردار ڪريو (پاڙيسري، رشتيدار، يا سنڀاليندڙ جيڪي توهان جي رهائش لاءِ فزيڪل اسپيئر ڪنجي رکن ٿا) ان ڳالهه جي تصديق ڪرڻ لاءِ ته ڪير توهان جي مدد ڪري سگهي ٿو جيڪڏهن لازمي نڪرڻ جو آرڊر جاري ڪيو وڃي يا عمارت جي لفٽ بند ٿي وڃي.
- Attach a waterproof pouch to your wheelchair or go bag containing a 14-day medication supply, printed prescriptions, equipment serial numbers, charging cords, service animal vaccination records, and a laminated Medical & Communication Passport.
طبي انحصار ايمرجنسي ٽريج فيصلي جو وڻ¶
When a power outage, severe weather warning, or mandatory evacuation impacts a household with durable medical equipment or functional access needs, follow this operational decision flowchart.
flowchart TD
Crisis["Disaster Warning, Power Grid Outage, or Evacuation Order"] --> EventType{"What is the primary immediate threat?"}
EventType -->|"Electrical Grid Collapse (Shelter in Place)"| PowerAudit["Switch DME to Internal Batteries & LiFePO4 Power Station<br/>Disable CPAP/Ventilator Heated Humidifiers (Saves 70% Power)<br/>Switch from Stationary O2 Concentrator to Pulse-Dose POC or E-Tanks"]
EventType -->|"Mandatory Building or Regional Evacuation"| EvacTriage["Activate Personal Support Network (3 Trusted Local Contacts)<br/>Deploy Track Evacuation Chair for Multi-Story Stairs (Never Elevators)<br/>Pack Manual Backup Wheelchair, Chargers, and 14-Day Meds"]
PowerAudit --> RuntimeCheck{"Will total battery or oxygen reserves fall below 4 hours before restoration?"}
RuntimeCheck -->|"Yes (Critical Life-Support Exhaustion)"| CallEMS["Call 911 / 112 / 999 Immediately<br/>Request Priority Transport to Hospital or Medical Special-Needs Shelter"]
RuntimeCheck -->|"No (Adequate 48-72h Reserve)"| ColdChain["Transfer Insulin & Biologics to Evaporative Frio Pouch or Cooler<br/>Monitor Indoor Heat/Cold Limits & Check In with Network Every 6 Hours"]
EvacTriage --> ShelterRight["Proceed to Accessible Public Shelter with Service Animal<br/>Enforce Legal Right to Keep Service Animal & DME Together"]
Danger
Never operate a portable gasoline, propane, or diesel generator inside a home, basement, attached garage, porch, or within 6 meters (20 feet) of any window, door, or vent to charge medical equipment. Generator exhaust contains odorless carbon monoxide (\(\text{CO}\)) gas at concentrations exceeding 10,000 parts per million, which can incapacitate and kill a medically fragile occupant in less than 10 minutes. For indoor backup medical power, use only zero-emission Lithium Iron Phosphate (\(\text{LiFePO}_4\)) battery power stations.
Durable Medical Equipment (DME) power engineering and watt-hour sizing¶
Surviving a 48-to-72-hour electrical blackout with life-sustaining medical devices requires calculating your exact Watt-Hour (\(\text{Wh}\)) load and optimizing how power is converted between the battery and the medical device.
How to calculate your required battery bank capacity¶
- Look at the manufacturer data plate on the bottom or back of your medical device (or its AC power brick) to find its power consumption in Watts (\(\text{W}\)). If the label lists only Volts (\(\text{V}\)) and Amperes (\(\text{A}\)), multiply them together (\(\text{Watts} = \text{Volts} \times \text{Amps}\)).
- Multiply the device's running wattage by the number of hours per day it must run to get Daily Watt-Hours (\(\text{Wh/day}\)).
- Divide by 0.85 to account for inverter and battery conversion overhead. For example, a ventilator drawing \(35\text{ W}\) for 24 hours consumes \(840\text{ Wh}\); divided by \(0.85\), you need a minimum \(1,000\text{ Wh}\) (1 kilowatt-hour) portable \(\text{LiFePO}_4\) power station per 24 hours.
| Durable Medical Equipment (DME) Device | Typical Running Power Draw (Watts) | 24-Hour Energy Requirement (Watt-Hours) | Critical Power-Conservation Engineering Tactics |
|---|---|---|---|
| Home Mechanical Ventilator (e.g., Trilogy, Astral, VOCSN) | 25 to 50 W (without external heated humidifier) | 600 to 1,200 Wh (Continuous 24-hour life support) | Keep both internal and detachable external click-in batteries fully charged (provides 6-14 hours built-in); replace electric heated-wire humidifier with passive Heat and Moisture Exchanger (HME) inline filters, saving 100+ W |
| Left Ventricular Assist Device (LVAD) (e.g., HeartMate 3) | 10 کان 18 W (هلندڙ)؛ 65 کان 100 W (بيٽري جو جوڙو چارج ڪرڻ) | 300 کان 500 وڇ | گھٽ ۾ گھٽ 4 کان 6 مڪمل طور تي چارج ٿيل 14V ليٿيم آئن مريض بيٽري پيڪ رکو (هر جوڙو 10-17 ڪلاڪ رهي ٿو)؛ موبائيل پاور يونٽ (MPU) کي خالص سائن ويو سان ڳنڍيل رکو \(\text{LiFePO}_4\) UPS |
| CPAP / BiPAP مشين (گرم هوادار ۽ ٽيوب سان ** آن**) | 75 کان 110 W | 600 کان 900 وڇ (في 8 ڪلاڪ ننڊ رات) | هائي ڊرين وارننگ: سڄي رات پاڻيءَ جي پليٽ گرم ڪرڻ سان توهان جي بيٽري جو 70 سيڪڙو ضايع ٿئي ٿو. بند ڪرڻ دوران فوري طور تي گرم نمي کي بند ڪريو |
| CPAP / BiPAP مشين (Humidifier OFF + DC-to-DC ڪيبل) | 12 کان 25 W | 100 کان 200 وڇ (في 8 ڪلاڪ ننڊ رات) | ٺاهيندڙ 12V/24V DC سگريٽ لائٽر ڪنورٽر ڪنڊ خريد ڪريو ۽ پنهنجي بيٽري اسٽيشن جي DC پورٽ ۾ پلگ ان ڪريو جڏهن ته AC انورٽر کي بند ڪريو (رن ٽائم کي وڌايو 3x کان 4x) |
| اسٽيشنري گهر آڪسيجن ڪنسنٽريٽر (5 کان 10 ايل / منٽ لڳاتار) | 300 کان 600 W (مسلسل ڪمپريسر موٽر) | 7,200 کان 14,400 ويھ | ننڍي بيٽري بئنڪن تي ڊگھي عرصي تائين نه ٿو هلائي سگھجي: فوري طور تي ڪمپريسڊ گيس اي-ٽينڪ/ ايم-ٽينڪ يا بيٽري پورٽيبل آڪسيجن ڪنسنٽريٽر (POC) پلس ڊوز موڊ تي |
| پورٽيبل آڪسيجن ڪنسنٽريٽر (POC) (پلس-دوز سيٽنگ 2-3) | 35 کان 90 W | 840 کان 2,160 وڇ | 12V DC ڪار پورٽ ذريعي چارج ڪريو يا شمسي سان مطابقت رکندڙ \(\text{LiFePO}_4\) اسٽيشن؛ گھٽ ۾ گھٽ ٻه ڊبل ظرف واري اسپيئر ليٿيم آئن بيٽرين کي پنھنجي کڻڻ واري بيگ ۾ رکو |
| ** Tracheostomy / Oral Airway Suction Pump** | 30 کان 60 W (واقعي 1-2 منٽ دفن) | 50 کان 100 وڇ | اليڪٽرڪ سکشن ناڪام ٿيڻ جي صورت ۾ هميشه هڪ دستي هٿ سان نچويل ميڪيڪل سکشن پمپ (جهڙوڪ Res-Q-Vac) پنهنجي گو بيگ ۾ جراثيم کان پاڪ سکشن ڪيٿيٽر سان گڏ رکو. |
| انٽرل فيڊنگ پمپ/IV انفيوژن پمپ (مثال طور، ڪنگارو جوئي) | 8 کان 15 واٽ | 150 کان 300 ويھ | اندروني بيٽري 12 کان 18 ڪلاڪ رهي ٿي؛ ھميشہ پيڪ ڪريو 60 mL انٽرل ڪيٿيٽر ٽِپ يا ENFit سرنجون ۽ ڪشش ثقل فيڊنگ بيگز دستي انجام ڏيڻ لاءِ گريوٽي بولس فيڊنگ بغير بجلي جي |
| خودڪار پيريٽونل ڊائلسز (APD) سائيڪلر | 120 کان 300 W | 1,000 کان 2,400 Wh (في 8-10 ڪلاڪ رات جو چڪر) | اليڪٽرڪ مشين سائيڪلر کان تبديل ڪرڻ لاءِ سامان ۽ تربيتي هدايتون هٿ تي رکو دستي مسلسل ايمبوليٽري پيريٽونل ڊائلسز (CAPD) ٽوئن بيگ گريوٽي ايڪسچينج |
| موٽر ٿيل پاور ويل چيئر / موبلٽي اسڪوٽر چارجر | 120 کان 250 W (4-8 ڪلاڪ ريچارج چڪر دوران) | 500 کان 1,200 Wh (في مڪمل ري چارج) | تيز رفتار موڊ کان بچڻ؛ رولنگ جي مزاحمت کي گھٽائڻ لاءِ نيوميٽڪ ٽائر فل ريٽيڊ PSI تي فليٽ ڪريو؛ مينهن جي نڪرڻ دوران ** جوائس اسٽڪ ڪنٽرولر** مٿان پنروڪ پلاسٽڪ جي ٿيلھي لپي |
ڇو DC-to-DC ڪيبلز ۽ \(\text{LiFePO}_4\) خالص سائن ويو انورٽرز اهم آهن¶
- اي سي انورٽر پاراسٽڪ نقصان کي ختم ڪرڻ: اڪثر طبي ڊوائيس (CPAPs، BiPAPs، پورٽبل آڪسيجن ڪنسنٽريٽر، ۽ فيڊنگ پمپ) اصل ۾ اندروني طور تي هلن ٿا 12V يا 24V Direct Current (DC) تي. جڏهن توهان ڊوائيس جي معياري وال ڪارڊ کي پورٽبل بيٽري اسٽيشن جي 120V/230V AC آئوٽليٽ ۾ پلگ ان ڪريو ٿا، بيٽري جي ڊي سي پاور کي اندروني انورٽر ذريعي AC ۾ ڦيرايو وڃي ٿو (جيڪو 10 کان 20 واٽ في ڪلاڪ جلندو آهي صرف آن رهي) ۽ پوءِ توهان جي ڊيوائس کي 5 سيڪڙو پاور اپ ڊي سي تائين واپس ڪيو ويو. بيٽري جي گنجائش جيئن گرمي. خريد ڪرڻ هڪ $25 کان $50 ٺاهيندڙ 12V/24V DC اڊاپٽر ڪيبل ۽ سڌو بيٽري اسٽيشن جي 12V ڪار ساکٽ ۾ پلگ ڪرڻ سان بجلي جي نقصان کي اڌ ۾ گھٽائي ٿو.
- Pure Sine-wave جي گهرج: حساس ميڊيڪل مائڪرو پروسيسرز، LVAD ڪنٽرولرز، ۽ گرم ٿيل نيوبولائزر وڌيڪ گرم ٿي سگهن ٿا يا خراب ٿي سگهن ٿا جيڪڏهن هڪ سستا ”تبديل ٿيل سائي-ويو“ انورٽر ۾ لڳايو وڃي. تصديق ڪريو ته ڪو به پورٽبل پاور اسٽيشن يا اڻ کٽ پاور سپلاءِ (UPS) جيڪو توهان خريد ڪيو آهي اهو بيان ڪري ٿو Pure Sine-wave AC output ۽ استعمال ڪري ٿو \(\text{LiFePO}_4\) (ليٿيم آئرن فاسفيٽ) بيٽري سيلز (جيڪي حرارتي طور تي مستحڪم، غير ٻرندڙ، ۽ 3، 0+ سائيڪل چارج لاءِ ريٽيڊ ٿيل آهن).
آڪسيجن تھراپي لچڪ ۽ ڪمپريسر سلنڈر جي مدت رياضي¶
جڏهن برقي طاقت ناڪام ٿئي ٿي، هڪ اسٽيشنري 5-کان-10-ليٽر-في-منٽ گهر آڪسيجن ڪنسنٽريٽر سيڪنڊن اندر آڪسيجن پيدا ڪرڻ بند ڪري ٿو، هڪ بلند آواز جي بيٽري الارم کي آواز ڏئي ٿو. هر آڪسيجن تي ڀاڙيندڙ مريض کي لازمي طور تي گهر ۾ ٺهيل گيس سلنرز (يا مائع آڪسيجن جي ذخيرن) کي برقرار رکڻ گهرجي ۽ اهو ڄاڻڻ گهرجي ته ڪئين ڪلاڪ آڪسيجن هڪ ٽينڪ ۾ رهي ٿو.
ٽينڪ جي مدت فارمولا ۽ معياري سلنڈر جي صلاحيت¶
ڳڻڻ لاءِ ته هڪ ڪمپريس ٿيل آڪسيجن سلنڈر ڪيترا منٽ لڳندو، معياري ڪلينيڪل سلنڈر استعمال ڪريو (\(F\)):
$\(\text{Duration in Minutes} = \frac{(\text{موجوده ٽينڪ گيج پريشر PSI ۾} - 200\text{PSI محفوظ بقايا}) \times \text{سائلڊر فيڪٽر } (F)}{\text{مقرر ڪيل وهڪري جي شرح ليٽر في منٽ ۾ (L/min)}}\)
- D-سائلڊر (\(F = 0.16\)، 2,000 psi تي ~350 کان 425 ليٽر رکي ٿو): 2 L/min مسلسل وهڪري تي، هڪ مڪمل 2,000 psi ڊي-ٽينڪ \(((2000 - 200) \times 0.16) \ma2{16/4{4 منٽ. ڪلاڪ)}}\).
- اي-سلينڊر (\(F = 0.28\)، 2,000 psi تي ~ 660 کان 680 ليٽر رکي ٿو): 2 L/min مسلسل وهڪري تي، هڪ مڪمل 2,000 psi E-ٽينڪ \(((2000 - 200) \times 0.28) / b2{2{2{5 منٽ. ڪلاڪ)}}\). 4 L/min تي، اهو صرف 2.1 ڪلاڪ رهي ٿو.
- **وڏي اسٽيشنري M-سلينڊر / H-سلينڊر (\(F = 1.56\text{ to }3.14\)، رکي ٿو 3,000 کان 7,000 ليٽر): ** بيڊ روم جي هڪ ڪنڊ ۾ سڌريل هڪ وڏو سائي اسٽيل H-ٽينڪ **50 کان 60 ڪلاڪ فراهم ڪري ٿو ** 50 کان 60 ڪلاڪ ** بغير ڪنهن به آڪسي / منٽ جي بجليءَ جي ** 50 کان 60 ڪلاڪن تائين.
ٽي غير ڳالهين واري آڪسيجن جي حفاظت جا ضابطا:
- رينچ ۽ اسپيئر بوڊوڪ واشر کي ٽيچر ڪريو: ڪيترائي اي-سلينڊر پن-انڊيڪس يوڪ ريگيوليٽر استعمال ڪندا آهن جن لاءِ سلٽ ٿيل دھات جي ضرورت هوندي آهي T-رينچ (ٽينڪ ڪي) والو جي اسٽيم کي کولڻ لاءِ ۽ هڪ رٻڙ/ڌاتو O-ring (Bodok seal) تيز پريشر جي ليڪ کي روڪڻ لاءِ. ٻه ڌاتو ٽينڪ جي چاٻيون ۽ ٽي اضافي او-رنگ سڌو سنئون ٽينڪ ڪارٽ تي نائلون ڪنڊ سان ڳنڍيو ته جيئن اونداهي ۾ سانس جي گھٽتائي دوران توهان ڪڏهن به رنچ جي ڳولا نه ڪريو.
- ڪنزروينگ ڊيوائسز (نيوميٽڪ ڊيمانڊ والوز): پنهنجي تنفس جي معالج کان پڇو ته ڇا توهان کي طبي طور تي صاف ڪيو ويو آهي ته توهان ايمرجنسي جي دوران توهان جي اي ٽينڪن تي ** نيوميٽڪ آڪسيجن جي حفاظت ڪندڙ ريگيوليٽر (ڊمانڊ والو)** استعمال ڪرڻ لاءِ. هڪ بچاءُ وارو ريگيوليٽر آڪسيجن جو هڪ بولس صرف ساهه کڻڻ جي پهرين اڌ سيڪنڊ دوران جاري ڪري ٿو بلڪه آڪسيجن ضايع ڪرڻ جي بجاءِ جڏهن توهان سانس ڪڍو ٿا، هڪ واحد E-ٽينڪ کي **4.2 ڪلاڪن کان 12 کان 14 ڪلاڪن تائين وڌايو.
- جسماني ۽ باهه جي حفاظت: ڪار جي سيٽ يا فرش تي ڪڏهن به دٻجي ويل آڪسيجن سلنڊر نه لڪايو جتي اوچتو بريڪ يا ٽِپ اوور پيتل جي والو جي ڳچيءَ کي ڪٽجي سگهي ٿو، 2,000-psi اسٽيل جي سلنڈر کي هڪ تيز رفتار تارپيڊو ۾ تبديل ڪري ٿو جيڪو ڀت ۾ گھڙڻ جي قابل ٿي سگھي ٿو. آڪسيجن کي گھٽ ۾ گھٽ 3 ميٽر (10 فوٽ) موم بتين، گيس اسٽو، اسپيس هيٽر ۽ پيٽروليم تي ٻڌل لپ بام کان پري رکو (خشڪ نٿن لاءِ پاڻي تي ٻڌل جيل استعمال ڪريو).
کولڊ چين دواسازي تحفظ ۽ 30-ڏينهن ريفيل بفر¶
انسولين جي حفاظت، GLP-1s، ۽ بايولوجڪ انجيڪشن بغير ريفريجريٽر جي¶
گھڻن ڏينهن جي بجلي جي بندش يا خارج ٿيڻ دوران، درجه حرارت حساس دوائن کي نظم و ضبط حرارتي انتظام جي ضرورت آهي:
- ** نه کليل بمقابله استعمال ۾ انسولين جا ضابطا:** انسولين جا نه کليل شيشيون ۽ اڳ ۾ ڀريل پينس 2.0 ۽ 8.0 درجا سينٽي گريڊ (36.0 ۽ 46.0 درجا فارن هائٽ) جي وچ ۾ محفوظ ٿيڻ گهرجن انهن جي اسٽيمپ ٿيل ختم ٿيڻ جي تاريخ تائين. بهرحال، جيڪڏهن ريفريجريشن مڪمل طور تي ناڪام ٿئي ٿي، ٻئي آمريڪي فوڊ اينڊ ڊرگ ايڊمنسٽريشن (FDA) ۽ ورلڊ هيلٿ آرگنائيزيشن تصديق ڪن ٿا ته سڀ کان وڌيڪ معياري انساني ۽ اينالاگ انسولين (باقاعده، اين پي ايڇ، گليرگائن، ليسپرو، اسپارٽ) ** مستحڪم ۽ ڪلينڪ طور استعمال لائق آهن ڪمري جي گرمي پد 25.0 کان 30.0 درجا Cel. فارنهائيٽ) تائين 28 ڏينهن تائين**. شيشي کي ريفريجريشن ۾ ڇڏي تاريخ کي نشان لڳايو ۽ بحران ختم ٿيڻ کان پوءِ ان کي تازو ريفريجريٽڊ اسٽاڪ سان تبديل ڪريو، جيتوڻيڪ انسولين 28 ڏينهن کان پوءِ به بوتل ۾ رهي.
- The absolute freezing Prohibition (0 C / 32 F): ** انسولين، ايپينفرائن، يا بايولوجڪ پين کي ڪڏهن به سڌو سنئون برف جي ڪعب، خشڪ برف، يا منجمد جيل برڪس جي خلاف نه رکو.** جيڪڏهن انسولين منجمد ٿئي ٿي، جيتوڻيڪ مختصر طور تي، برف جي ڪرسٽل ان جي نازڪ چوٿون پروٽين جي جوڙجڪ کي ڇڪيندا آهن. اڳي منجمد ٿيل انسولين ۾ گلوڪوز گھٽ ڪرڻ واري سرگرمي آهي ۽ زندگي جي خطري واري ** ذيابيطڪ ڪيٽوآسيڊسس (DKA)** کي شروع ڪري سگهي ٿي. دوائن جي دٻن کي هميشه ٿلهي هٿ جي توليه ۾ ويڙهيو يا انهن کي پلاسٽڪ فوڊ ڪنٽينر جي اندر رکو جيڪو آئس پيڪن کان الڳ ٿيل آهي ڪولر اندر، ۽ منٽ / وڌ ۾ وڌ فرج ٿرماميٽر سان مانيٽر ڪريو.
- هاٽ جي لهرن لاءِ بخاري کولنگ پائوچز (فريو بيگز): گرم موسمن ۾ يا اونهاري جي بليڪ آئوٽ دوران برف کان سواءِ، اندر اندر انسولين ۽ ايپينفرائن پين کي ذخيرو ڪريو پوليڪريلامائيڊ ڪرسٽل ايپوريٽو پائوچز (فريو بيگز). 10 منٽن لاءِ ٿڌي نل جي پاڻيءَ ۾ اندرين ٿلهي کي ٻرڻ سان ڪرسٽل فعال ٿي ويندا آهن. قدرتي پاڻي جو بخار اندرون محفوظ رکي ٿو 18 کان 26 ڊگري سينٽي گريڊ (64 کان 79 F) لاءِ 45 لڳاتار ڪلاڪ جيتوڻيڪ جڏهن ٻاهران هوا 38 درجا سينٽي گريڊ (100 F) هجي، ۽ ڪنهن به صاف پاڻي سان لامحدود طور تي ري چارج ٿي سگهي ٿي.
14 کان 30 ڏينهن جي ايمرجنسي نسخي واري رزرو جي تعمير¶
آفت کان اڳ انشورنس "ريفيل-تو-جلد" بلاڪ کي ختم ڪرڻ لاء:
- Switch maintenance drugs to 90-day fills: Ask your physician to rewrite your 30-day maintenance prescriptions (for antihypertensives, anticonvulsants, inhalers, thyroid hormones, and immunosuppressants) as 90-day mail-order or retail fills, which allows you to refill at day 68 to 75 and immediately creates a rolling 15-to-20-day emergency reserve.
- Invoke State of Emergency pharmacy override laws: More than 30 US states, Canadian provinces, and the UK NHS ("Emergency Supply at Patient Request") have statutory rules requiring pharmacies and insurers to dispense an emergency 7-to-30-day supply of non-controlled maintenance medications without a new physician signature whenever the governor, premier, or public health authority declares a State of Emergency or when a patient is evacuated from their home pharmacy. Carry a photographed copy of all prescription bottle labels (showing the Rx number, pharmacy phone number, drug name, milligram dose, and National Drug Code / NDC) on your phone and in your waterproof go-bag folder.
Renal dialysis, enteral nutrition, and ostomy crisis protocols¶
Hemodialysis patients: the "3-Day Emergency Diet"¶
During major hurricanes, floods, or earthquakes, outpatient hemodialysis clinics may lose municipal water treatment pressure or electrical power for 48 to 72 hours. Missing two consecutive hemodialysis sessions causes two immediate life threats: lethal hyperkalemia (high blood potassium causing sudden cardiac arrest) and acute pulmonary edema (fluid overload drowning the lungs).
Every hemodialysis patient must keep a 3-Day Renal Emergency Kit at home and switch immediately to the National Kidney Foundation (NKF) / KCE 3-Day Emergency Renal Diet the moment a disaster threatens clinic closures:
- Strict fluid restriction: Limit total liquid intake (including water, ice chips, soup, gelatin, and gravy) to 500 to 700 milliliters (16 to 24 fluid ounces) per 24 hours (plus urine output if you still produce urine). Weigh yourself on a bathroom scale every morning; gaining more than 1.5 to 2.0 kilograms (3.3 to 4.4 pounds) between treatments requires urgent medical contact.
- Strict low-potassium and salt-free food selection: Eliminate all high-potassium foods (bananas, oranges, potatoes, tomatoes, avocados, chocolate, beans, nuts, milk, and salt substitutes containing potassium chloride). Stock shelf-stable low-potassium, unsalted foods: canned tuna or chicken rinsed in water, white bread or unsalted crackers, white rice, applesauce, canned peaches or pears (drain the heavy syrup), hard candies, honey, and marshmellows for high-calorie energy.
- Emergency potassium-binding medication: Ask your nephrologist ahead of time whether you should keep a prescribed emergency dose of an oral potassium binder (such as sodium zirconium cyclosilicate [Lokelma], patiromer [Veltassa], or sodium polystyrene sulfonate [Kayexalate]) inside your home emergency kit, along with your typed Dialysis Treatment Prescription Sheet (listing your dialyzer model, blood flow rate, dialysate bath potassium/calcium concentration, heparin dose, vascular access type, and target dry weight) so any backup hospital or out-of-town clinic can dialyze you immediately.
- Vascular access hemorrhage kit: Keep two sterile gauze packs and a bottle of povidone-iodine taped to your go bag. If an arteriovenous (AV) fistula or graft ever bleeds during an emergency, apply firm two-finger pinpoint direct pressure (or press a clean bottle cap directly over the bleeding pinhole for 10 to 15 minutes) while calling 911 / 112 / 999, never wrap a tight tourniquet or blood pressure cuff above a bleeding dialysis fistula, which increases venous back-pressure and makes it bleed harder.
Sensory, cognitive, and neurodivergent emergency adaptations¶
Standard emergency broadcasts and chaotic public shelters can be inaccessible or overwhelming for deaf, blind, non-speaking, autistic, or dementia-affected individuals. Configure your household kit using this functional adaptation matrix.
| Functional / Sensory Need | Specialized Alert & Preparedness Equipment | Evacuation & Shelter Operational Protocol |
|---|---|---|
| Deaf or Hard of Hearing (DHH) | 177-candela xenon strobe smoke/CO alarms; NOAA weather radio with under-pillow vibrating bed-shaker puck | Pack a 30-day supply of zinc-air hearing aid batteries (keep dry with silica desiccant packets) or cochlear implant USB power bank; attach a waterproof notepad + thick black markers to your belt to communicate with rescuers wearing masks |
| انڌو يا گهٽ بصارت | بريل/اٿاريل ڊاٽ ٽيڪٽائل ليبل ۽ دوائن جي بوتلن ۽ ايمرجنسي فوڊ جين تي تيز ڪنٽراسٽ پيلو/ڪارو ٽيپ | اسٽيج ** ٻه اضافي فولڊنگ سفيد موبليٽي ڪينز** (هڪ پلنگ وٽ، هڪ اڳيان دروازي وٽ) جيئن ٽٽل شيشي ۽ ملبہ اڪثر ڪري پرائمري ڪنڊن کي ٽوڙيندا آهن؛ ڪٽ جي مزاحمتي ڪيولر/چمڙي جي ڪم جا دستانا پنهنجي کيسي ۾ رکو ته جيئن ڀتين ۽ ڏاڪڻين کي ڇڪڻ سان محفوظ طور تي صاف ڪرڻ لاءِ |
| غير ڳالهائڻ وارا / AAC استعمال ڪندڙ (ALS، سيريبرل پالسي، اسٽروڪ، آٽزم) | لامحدود ** لو-ٽيڪ تصوير / الفابيٽ ڪميونيڪيشن بورڊ** (ڪڏهن به صرف هڪ iPad تي ڀروسو نه ڪريو جنهن جي بيٽري ختم ٿي سگهي) | هڪ لامحدود ميڊيڪل ڪميونيڪيشن پاسپورٽ کي پنهنجي سيني تي پن ڪريو؛ اسمارٽ فون تي آف لائن Text-to-Speech ائپس انسٽال ڪريو، جيڪو بيٺو بلوٽوٿ اسپيڪر ۽ 20,000 mAh پاور بئنڪ سان جوڙيو ويو آهي |
| آٽزم اسپيڪٽرم ۽ حسي پروسيسنگ | اوور-ڪن 25-30 dB غير فعال/ فعال شور منسوخ ڪرڻ وارا هيڊفون، گندي لفافي جي چوڌاري چشما، غير خوشبودار وائپس | هڪ ** حسي سينچري بيگ ** پيڪ ڪريو: واقف وزن وارو ليپ پيڊ يا ڪمپريشن ويسٽ، ترجيحي شيلف-مستحڪم محفوظ کاڌ خوراڪ (ڊائي هائيڊريشن/بُک کان بچڻ لاءِ کاڌي جي بناوت کان بچاءُ لاءِ)، ۽ روشن، شور واري عوامي پناهگاهن لاءِ هڪ ننڍڙو پاپ اپ رازداري خيمو |
| الزائمر جي بيماري، ڊيمينشيا ۽ ٽي بي آئي | پنروڪ GPS سيلولر/آر ايف لوڪيٽر واچ يا سلائي ان ايئر ٽيگ؛ MedicAlert نقاشي ٿيل ID ڪنگڻ جنهن کي آسانيءَ سان بند نه ڪري سگهجي | افراتفري جي نڪرڻ واري پيڪنگ دوران ماڻهوءَ کي ڪڏهن به بي پرواهه نه ڇڏي ڏيو (وڏي گھمڻ/ ڀڄڻ جو محرڪ)؛ پرسڪون، هڪ قدم کنڪريٽ جملن ۾ ڳالهايو؛ هڪ آرامده واقف ڪمبل يا فوٽو البم ۽ هڪ موجوده مڪمل ڊگھائي تصوير پيڪ ڪريو |
ملٽي-ڪهاڻي ڏاڪڻ جي نڪرڻ ۽ ويل چيئر جي لچڪ¶
عمارت جي باهه، زلزلي، يا بليڪ آئوٽ دوران، لفٽ خود بخود گرائونڊ فلور تي ياد اچي ٿي يا طاقت وڃائي ٿي. ڪڏهن به قبضي واري 150-ڪلوگرام (330-پائونڊ) موٽرسائيڪل پاور ويل چيئر کي ڏاڪڻ جي اڏام هيٺ کڻڻ جي ڪوشش نه ڪريو، جنهن ۾ سوار ۽ بچاءُ ڪندڙ ٻنهي لاءِ موتمار ڦاٽڻ جو خطرو آهي.
- فائر ريڊ ٿيل ڏاڪڻين ۾ بچاءُ جي مدد جو علائقو (ريفيوجي ايريا): جديد ملٽي اسٽوري ڪمرشل ۽ رهائشي عمارتن ۾ فائر ريٽيڊ بند ٿيل ڏاڪڻيون آهن جن ۾ هڪ مقرر ٿيل لينڊنگ اسپيس سان ليس آهي ٻه طرفي انٽرڪام فائر ڪمانڊ پينل سان ڳنڍيل آهي. جيڪڏهن توهان مٿين منزل تي آهيو، توهان جي سپورٽ ٽيم يا فائر فائائيندڙن جي انتظار ۾، فوري طور تي دٻاء واري ڏاڪڻ جي لينڊنگ ۾ (توهان جي پويان وڏي باهه جو دروازو بند ڪرڻ) ڏانهن وڃو.
- ** ٽريڪ Evacuation Chairs (مثال طور، Evac+Chair, Stryker Stair-PRO):** جيڪڏھن توھان رھو ٿا يا گرائونڊ فلور کان مٿي ڪم ڪريو ٿا، توھان جي بلڊنگ مينيجمينٽ يا گھر مالڪن جي ايسوسيئيشن سان صلاح ڪريو ته ** ٽريڪ ٿيل ڏاڪڻ خالي ڪرڻ واري ڪرسي ** مٿي لينڊنگ تي. ڪرسيءَ جي هيٺان رٻڙ واري رگڻ واري بيلٽ، ڏاڪڻ جي نانگن کي پڪڙي ٿي، جنهن جي ڪري هڪ 60-ڪلوگرام سنڀاليندڙ هڪ 180-ڪلوگرام مسافر کي بغير کڻڻ جي آسانيءَ سان ڏاڪڻ تان هيٺ لهي سگهي ٿو.
- ٻه ماڻهو ”اڳتي ۽ پويان“ (Extremity) Carry & Webbing Slings: جيڪڏهن ڪا ٽريڪ چيئر موجود نه آهي ۽ توهان کي گهرجي ته وهيل چيئر استعمال ڪندڙ کي فوري طور تي ڏاڪڻ تان هيٺ لاٿو:
- پنهنجي اپارٽمنٽ جي دروازي جي اڳيان هڪ هيوي ڊيوٽي ڪئنوس جي منتقلي واري سلنگ کي چار پيڊل ٿيل هينڊلز (يا مضبوط ٽنگن سان هڪ مضبوط ڊائننگ چيئر) رکو.
- سامان کان سواءِ ٻه فردن جي اڳيان ۽ پٺيان ڪيري لاءِ: ماڻهوءَ کي ويل چيئر کان ٻاهر فرش يا هڪ قدم تي منتقل ڪريو. بچائيندڙ 1 (مضبوط لفٽر) ماڻهوءَ جي پٺيءَ پٺيان گوڏا کوڙي، ماڻهوءَ جي ٻنهي بغلن جي هيٺان پهچندو آهي، ۽ ماڻهوءَ جي ڪراس ڪيل کلائي کي مضبوطيءَ سان ماڻهوءَ جي سيني ۾ پڪڙيندو آهي (مريض جي ننگي اسٽريم تي هٿ ڪڏهن به بند نه ڪريو، ۽ اها گرفت صرف ان صورت ۾ استعمال ڪريو، جڏهن ماڻهوءَ کي مٿي تي ڪا تڪليف نه هجي) بچائيندڙ 2 ماڻهوءَ جي پيرن جي وچ ۾ يا ان جي ڀرسان بيٺو آهي، جن کي ڏاڪڻ کان هيٺ منهن ڪيو وڃي ٿو ۽ ٻنهي گوڏن/رانن هيٺان پينگهه. Rescuer 1 جي ڳڻپ تي، ٻئي هڪ ئي وقت پنهنجن ٽنگن سان گڏ کڻندا آهن (پٺيون سڌيون) ۽ هڪ دفعي هڪ ڏاڪڻ تان هيٺ لهي ويندا آهن.
- دستي ويل چيئر ايمرجنسي مينٽيننس ڪٽ: پنهنجي مينوئل يا پاور ويل چيئر سيٽ جي هيٺان هڪ ننڍڙو زپ ٿيل پائوچ ڳنڍيو جنهن ۾: (a) هڪ هيوي ڊيوٽي ٽائر جي مرمت جي کٽ، اسپيئر انر ٽيوب، يا سولڊ فليٽ فري پولي يوريٿين ٽائر انسرٽس ان سان گڏ هڪ ڪمپيڪٽ CO2/هٿ سان پمپ ٿيل اسٽريٽ سان گڏ آهن. ۽ ٽٽل گلاس)؛ (b) آلن هيڪس ڪيز ۽ 10mm/13mm رينچ سان گڏ هڪ گھڻائي وارو اوزار؛ © هڪ ڀريل چمڙي جو جوڙو پشنگ دستانن؛ ۽ (d) ڇپيل تصويري هدايتون مددگارن کي ڏيکاريندي ته ڪيئن توهان جي پاور ويل چيئر موٽرز تي ٻه مڪينيڪل فري ويل ڪلچ ليورز کي فلپ ڪيو وڃي ته جيئن بيٽري ڊهي وڃي ته ڪرسي کي دستي طور تي دٻائي سگهجي.
قانوني پناهه جا حق، خدمت جانورن، ۽ افاديت جي ترجيحي رجسٽري¶
- سروس جانورن لاءِ غير مشروط قانوني تحفظ: تحت آمريڪن سان گڏ معذوري ايڪٽ (ADA ٽائيٽل II ۽ ٽائيٽل III) ۽ فيما اسٽافڊ ايڪٽ پاليسي 104-009-2 آمريڪا ۾، برابري ايڪٽ 2010 برطانيه ۾، **ڪناڊا ۾ **Accessible/Accessible Humandes" EU جي شهري حقن جي برابر هدايتون، سڀني عوامي ايمرجنسي پناهگاهون، ريڊ ڪراس پناهه گاهون، خالي ڪرڻ واريون بسون، ۽ عارضي ڊزاسٽر هائوسنگ کي لازمي طور تي معذور ماڻهن کي انهن جي خدمت جانورن سان گڏ داخل ڪرڻ گهرجي. هڪ شيلٽر مئنيجر **قانوني طور تي توهان کي موڙي نه ٿو سگهي، توهان کي ڌار ڌار ڪمري ۾ رکي، يا توهان کان هڪ تربيت يافته خدمت ڪتي کي ٻاهرين پالتو جانورن جي ڪتن ۾ رکڻ جي ضرورت آهي، جيتوڻيڪ ٻيا پناهه گاهه ڪندڙ ڪتن جي الرجي يا ڪتن جي خوف جي دعويٰ ڪن ٿا (پناه گاهه کي ٻنهي جي ويجهڙائيءَ ۾ جاءِ ڏيڻ گهرجي هاٿي جي سامهون ڪنڊن تي فاصلو رکڻ سان). هڪ 7-ڏينهن جي سروس ڊاگ ڪٽ کي پيڪ ڪريو جنهن ۾ کاڌو، ڪلاپيبل واٽر باؤل، ريبيز سرٽيفڪيٽ، اسپيئر ليش/هارنس، ۽ 4 ربر ٿيل ڪتي جا بوٽ توهان جي ڪتي جي پنن جي پيڊن کي ٽٽل شيشي، گرم اسفالٽ ۽ ڪيميائي ٻوڏ جي پاڻي کان بچائڻ لاءِ.
- اڄ ئي اليڪٽرڪ يوٽيلٽي ”ميڊيڪل بيس لائين“ ۽ HHS emPOWER رجسٽرين ۾ داخل ٿيو:
- يوٽيلٽي ميڊيڪل پرائيرٽي پروگرام: اڄ ئي پنهنجي اليڪٽرڪ ۽ گيس يوٽيلٽي ڪمپنين کي ڪال ڪريو ۽ انهن جي درخواست ڪريو ميڊيڪل بيس لائين/ لائف سپورٽ آلات سرٽيفڪيشن فارم (توهان جي ڊاڪٽر يا PA/NP پاران دستخط ٿيل). هڪ دفعو توهان جي يوٽيلٽي جي لائف سپورٽ رجسٽري تي رجسٽر ٿيڻ کان پوءِ، يوٽيلٽي کي قانوني طور تي گھربل آهي گھڻن دائرن ۾ توهان کي اڳواٽ در در تائين يا پاڻمرادو وارننگ ڏيڻ لاءِ منصوبابندي ڪيل پبلڪ سيفٽي پاور شٽ آفز (PSPS) کان اڳ، توهان جي ميٽر کي ريموٽ ڪنيڪشن کان ڇڏايو، ۽ توهان جي پاڙيسري ٽرانسفارمر سرڪٽ کي ترجيحي بحالي واري لائن لاءِ پرچم لڳايو.
- ميونسپل اسپيشل-نيڊس ۽ اسمارٽ 911 رجسٽري: پنھنجي گھر کي رجسٽر ڪريو پنھنجي مقامي شھر يا ڪائونٽي سان ايمرجنسي مئنيجمينٽ فنڪشنل نيڊس رجسٽري ۽ ٺاھيو مفت Smart911 پروفائل (
smart911.comآمريڪا ۾) ته جيئن توهان 911 ڊائل ڪريو ٿا، ڊسپيچر جي اسڪرين خود بخود ڏيکاري ٿي توهان جي صحيح منزل جي سطح، طاقت تي منحصر وينٽيليٽر يا آڪسيجن جي حالت، متحرڪ سامان، ۽ مواصلاتي هدايتون.
Tip
هڪ ٽي ماڻهون پرسنل سپورٽ نيٽ ورڪ قائم ڪريو (توهان جي گهر جي 5 منٽن اندر ڀروسي وارا پاڙيسري، مائٽن، يا عمارت سازي جو عملو) جيڪي توهان جي سامهون واري دروازي وٽ فزيڪل اسپيئر چاٻي يا سمارٽ لاڪ ڪوڊ رکن ٿا، ڄاڻو ته توهان جي پاور ويل چيئر تي مينوئل فري ويل ليورز کي ڪيئن ختم ڪجي، ۽ ڄاڻو ته توهان جو بيڪ اپ آڪسيجن بيگز ڪٿي وڃي رهيو آهي.
اڪثر پڇيا ويا سوال¶
How can I run my CPAP or BiPAP machine for multiple nights during a power outage without draining my backup battery?
Two simple adjustments will triple or quadruple your CPAP or BiPAP battery runtime during a blackout. First, go into your machine's settings menu and **turn the heated humidifier and heated tubing completely OFF** (you can still leave distilled water in the chamber for passive pass-over humidity); the heating plate consumes 65 to 75 percent of the machine's total electricity. Second, do not plug your CPAP's standard AC wall cord into the 120V/230V AC outlet of your portable battery station. Instead, buy the manufacturer's **12V or 24V DC cigarette-lighter converter cable** (available for ResMed, Philips, and Luna units) and plug it directly into the battery station's DC port with the AC inverter turned off. Together, these two steps drop nightly consumption from 800 Wh down to just 100 to 150 Wh, allowing a single 500 Wh LiFePO4 battery to run your CPAP for 3 to 4 full nights.
What should I do if my home electric oxygen concentrator stops working when the power goes out?
Immediately switch your nasal cannula tubing from the silenced electric concentrator over to your backup **compressed-gas oxygen cylinder (E-tank or M-tank)** or your battery-powered **Portable Oxygen Concentrator (POC)**. Open the cylinder valve slowly counter-clockwise using the attached tank wrench until the pressure gauge reads tank PSI, then dial the flowmeter knob to your prescribed liters per minute. Next, calculate your remaining hours of oxygen: a full 2,000 psi E-tank provides roughly **4.2 hours** of continuous oxygen at 2 L/min (or 2.1 hours at 4 L/min). If you use a pulse-dose conserving regulator, that same tank lasts 12+ hours. If your total backup oxygen supply will drop below 2 to 4 hours before power is restored, call your oxygen supplier's 24-hour emergency line or dial 911 / 112 / 999 for transport to an energized medical shelter or hospital.
How long does insulin stay safe if my refrigerator loses power during a summer heatwave?
According to the US Food and Drug Administration (FDA) and manufacturer stability testing, both opened and unopened vials or pens of most human and analog insulins remain clinically potent at room temperatures between **15 and 30 degrees Celsius (59 and 86 degrees Fahrenheit) for up to 28 days**. However, insulin rapidly loses potency if exposed to extreme heat above **30 to 35 degrees Celsius (86 to 95 F)** or direct sunlight, and it is **permanently destroyed if it freezes (0 C / 32 F)**. During a summer blackout, store your insulin inside a water-activated **evaporative cooling pouch (such as a Frio bag)**, which keeps insulin safely around 18 to 26 C (64 to 79 F) for 45 hours using only tap water, or inside an insulated cooler with chilled gel packs wrapped in a thick towel so the insulin vial never touches ice directly.
Can a public emergency shelter or evacuation bus turn away my service dog because of other people's allergies?
No. Under Title II and Title III of the Americans with Disabilities Act (ADA), the federal Stafford Act in the United States, the Equality Act 2010 in the United Kingdom, and international disability protection laws, public emergency shelters, Red Cross facilities, and disaster evacuation buses are legally required to admit individuals with disabilities alongside their service animals into general population areas. Shelter staff **cannot** ask for "certification papers" (in the US, they may only ask two questions: (1) Is the dog a service animal required because of a disability? and (2) What work or task has the dog been trained to perform?), cannot isolate you, and cannot force your service dog into a pet kennel. If another shelter resident has a severe dog allergy, the shelter must accommodate both of you by assigning cots in separate areas of the shelter.
What should a kidney hemodialysis patient do if a storm or disaster closes their dialysis clinic for 2 to 3 days?
If a disaster interrupts your scheduled hemodialysis sessions, immediately start your **3-Day Emergency Renal Diet** to prevent fatal potassium overload (hyperkalemia) and lung fluid congestion (pulmonary edema). Restrict all fluid intake (water, ice, soup) to no more than **500 to 700 milliliters (2 to 3 cups) per day**, eliminate all salt and high-potassium foods (bananas, oranges, potatoes, tomatoes, beans, milk, nuts, and chocolate), and eat only low-potassium, salt-free shelf-stable foods (rinsed canned tuna/chicken, white bread, white rice, applesauce, drained canned peaches, and hard candies). If your nephrologist has prescribed an emergency potassium-binding powder (such as Lokelma, Veltassa, or Kayexalate), take it as directed on your emergency plan, and call 911 / 112 / 999 immediately if you develop chest tightness, shortness of breath while lying flat, or severe leg weakness.
How do I push a heavy electric power wheelchair if the battery dies during an evacuation?
When an electric power wheelchair is in normal driving mode, electromagnetic motor brakes lock the drive wheels automatically whenever the joystick is centered so the chair cannot roll down hills. If the battery dies and you try to push the chair without disengaging those brakes, the wheels will remain locked solid. Look at the two main drive motors located near the rear or center drive wheels: every power wheelchair has **two mechanical "Freewheel" (Neutral/Drive) levers** (usually marked with a picture of an unlocked padlock or a person pushing a chair). Flip both levers from "Drive" to "Neutral/Freewheel" to mechanically disengage the motor gearboxes so a caregiver can push the chair manually on flat ground (never flip a power chair into freewheel while sitting on a steep ramp or hill, as you will lose all braking control).
لاڳاپيل ھدايتون¶
- Family emergency communication and evacuation plan
- ايمرجنسي وڃڻ وارو بيگ ۽ 72 ڪلاڪ پيڪ
- گھر جي پهرين امداد سپلائي کٽ چيڪ لسٽ
- Power outage emergency heating and food preservation
- تاريخي گرمي گنبد جي بقا
- Emergency numbers and international dispatch
ذريعن جي چڪاس ڪئي وئي¶
- فيڊرل ايمرجنسي مئنيجمينٽ ايجنسي (FEMA) ۽ آمريڪي ريڊ ڪراس (آمريڪا): معذور ماڻهن ۽ ٻين خاص ضرورتن لاءِ آفت جي تياري، ۽ FEMA پاليسي 104-009-2 (ڊزاسٽر آپريشنز ۾ معذور ماڻهن کي شامل ڪرڻ).
- يو ايس ڊپارٽمينٽ آف هيلٿ اينڊ هيومن سروسز (HHS, United States): HHS emPOWER Program (Electricity-Dependent Medicare Beneficiary Durable Medical Equipment Guidelines) ۽ ايمرجنسي ۾ انسولين اسٽوريج تي FDA گائيڊنس.
- يو ايس ڊپارٽمينٽ آف جسٽس سول رائٽس ڊويزن (آمريڪا): رياست ۽ مقامي حڪومتن لاءِ ADA بهترين عملن جي ٽول کٽ، باب 7، ايمرجنسي مئنيجمينٽ ۽ سروس اينيملز ان شيلٽرز.
- سينٽرز فار ميڊيڪيئر ۽ ميڊيڪيڊ سروسز (سي ايم ايس) ڪڊني ڪميونٽي ايمرجنسي رسپانس (KCER) ۽ نيشنل ڪڊني فائونڊيشن (NKF، آمريڪا): هنگامي حالتن جي تياري، ڊائليسس تي ماڻهن لاءِ هڪ گائيڊ (3-ڏينهن ايمرجنسي رينل ڊيٽ).
- ورلڊ هيلٿ آرگنائيزيشن (ڊبليو ايڇ او) ۽ گڏيل قومن جو دفتر آفت جي خطري جي گھٽتائي لاءِ (UNDRR، بين الاقوامي): معذوريءَ تي ھدايتون شامل آھن آفت جي خطري جي گھٽتائي ۽ مددگار ٽيڪنالاجي تسلسل.
- نيشنل هيلٿ سروس (اين ايڇ ايس) ۽ برطانيه جي ڪابينا آفيس (برطانيه): ڪلينڪ طور تي ڪمزور فردن ۽ ترجيحي خدمتن جي رجسٽرڊ (PSR) معيارن لاءِ هنگامي تياري تي هدايت.